The Hermle Z 326 K refrigerated benchtop centrifuge with 18 000 1/min speed and 23 545 x g max. RCF value, 4 x 100 ml capacity. Temperature from -20 °C to 40 °C, CFC-free refrigeration system, in a width of only 40 cm.
- Brand:
- Hermle
- Product code:
- DI023
Hermle Z 326 K refrigerated centrifuge
The Hermle Z 326 K refrigerated benchtop centrifuge, with its wide range of rotors, is suitable for a variety of research applications. Switching between swing-out, micro and large-volume angle rotors can be done simply, within seconds. The refrigeration system is located at the back of the unit, so it requires only 40 cm of width on the lab bench.
Technical data
- Max. speed: 18 000 1/min
- Max. RCF: 23 545 x g
- Max. capacity: 4 x 100 ml
- Speed range: 200 - 18 000 1/min
- Temperature range: -20 °C - 40 °C, in 1 °C steps
- Run time: 99 h 59 min / 59 min 50 s, in 1 min / 10 s steps
- Dimensions (W x H x D): 40 cm x 36 cm x 70 cm
- Weight: 60 kg
- Power consumption: 660 W
- Order numbers: 311.00 V01 (230 V / 50-60 Hz), 311.00 V02 (120 V / 50-60 Hz)
Features
- Maintenance-free induction drive
- Microprocessor control with large LCD display
- Motorized lid lock
- Automatic rotor recognition with overspeed protection, immediately after inserting the rotor
- Active, rotor-specific imbalance detection and shut-off, imbalance-tolerant drive
- CFC-free refrigeration system, pre-cooling program, and cooling that operates even at standstill
- Display of sample temperature
- One-hand operable, splash-proof digital display, usable even with protective gloves
- Speed setting in 1/min and g-value, in steps of 10
- 10 acceleration and 10 braking levels, with the option of coasting to a stop without braking
- Storage of up to 99 programs together with the rotor type
- Run time from 10 s to 99 h 59 min, or continuous operation
- Quick button for short spins, diagnostic program
- Acoustic signal at the end of the run, several melodies selectable
- Noise level below 60 dBA at max. speed
- Manufactured in accordance with international safety regulations (IEC 61010)
Rotors
| Rotor | Order number | Max. speed | Max. RCF | Max. radius | Temperature at max. speed |
|---|---|---|---|---|---|
| Swing-out rotor, 4-place (0-90°) | 220.72 V06 | 5 000 1/min | 3 885 x g | 13,9 cm | -3 °C |
| Swing-out rotor, 4-place (0-90°) | 221.12 V03 | 4 500 1/min | 3 350 x g | 14,8 cm | -7 °C |
| Swing-out rotor, 2 x 3 microtiter plates (0-90°) | 221.16 V03 | 4 500 1/min | 2 739 x g | 12,1 cm | -6 °C |
| Angle rotor 30 x 15 ml (35°) | 221.19 V02 | 4 500 1/min | 2 829 / 2 467 x g | 12,5 / 10,9 cm | -9 °C |
| Angle rotor 12 x 15 ml (28°) | 220.96 V04 | 6 000 1/min | 4 185 x g | 10,4 cm | -7 °C |
| Angle rotor 6 x 50 ml (28°) | 220.97 V04 | 6 000 1/min | 3 823 x g | 9,5 cm | -8 °C |
| Angle rotor 6 x 80 ml (25°) | 220.78 V05 | 11 000 / 13 500 1/min | 13 933 / 20 986 x g | 10,3 cm | 17 °C |
| Angle rotor 6 x 85 ml (38°) | 221.18 V02 | 9 000 1/min | 10 414 x g | 11,5 cm | -5 °C |
| Angle rotor 4 x 85 ml (30°) | 221.20 V02 | 12 000 1/min | 14 811 x g | 9,2 cm | 5 °C |
| Angle rotor 6 x 50 ml (26°) | 221.22 V02 | 12 000 1/min | 13 523 x g | 8,4 cm | -5 °C |
| Angle rotor 20 x 10 ml (30°) | 221.28 V02 | 12 000 1/min | 15 938 x g | 9,9 cm | 2 °C |
| Angle rotor 30 x 1,5 / 2,0 ml (45°, hermetically sealed) | 221.17 V03 | 13 000 / 14 000 1/min | 17 760 / 20 598 x g | 9,4 cm | 9 °C |
| Angle rotor 24 x 1,5 / 2,0 ml (45°) | 220.87 V09 | 14 000 / 15 000 1/min | 18 625 / 21 381 x g | 8,5 cm | 6 °C |
| Angle rotor 24 x 1,5 / 2,0 ml (45°, hermetically sealed) | 220.87 V10 | 14 000 / 15 000 1/min | 18 625 / 21 381 x g | 8,5 cm | 6 °C |
| Angle rotor 12 x 1,5 / 2,0 ml (40°) | 221.23 V02 | 18 000 1/min | 23 545 x g | 6,5 cm | -2 °C |
| Angle rotor 44 x 1,5 / 2,0 ml (40°) | 220.88 V07 | 13 500 1/min | 17 115 / 14 466 x g | 8,4 / 7,1 cm | 6 °C |
| Angle rotor 64 x 0,4 ml (40°) | 220.92 V06 | 13 500 1/min | 16 708 / 14 059 x g | 8,2 / 6,9 cm | 2 °C |
| Angle rotor 4 x 8-well PCR strip (45°) | 221.38 V01 | 15 000 1/min | 15 344 x g | 6,1 cm | -1 °C |